06 May

Tracheostomy Ventilation: METHODS

All of the vital capacities (VCs) of the patients were measured in both sitting and supine positions. Any patients for whom food aspiration was suspected had methylene blue tests. Two patients with positive methylene blue tests had high positioned tracheosto­mies and were converted to noninvasive alternatives of ventilatory support without long-term cuff deflation. Two other patients had tracheoesophageal fistulas that were repaired before further atten­tion could be paid to their tracheostomy tubes.

Continue Reading »

05 May

Tracheostomy Ventilation: PATIENTS

One hundred four long-term ventilator-dependent patients with neuromuscular respirator) insufficiency were referred for pulmo­nary rehabilitation. Seventy-eight of these patients were dependent on TIPPV with inflated cuffs at referral. The other 26 patients were converted to TIPPV after varying periods of time receiving nonin­vasive methods of assisted ventilation. The 104 patients presented with the following diagnoses: traumatic high level quadriplegia in 38 patients, postpolio in 22 patients, Duchenne muscular dystrophy in 19 patients, non-Duchenne myopathies in 14 patients, polymyo­sitis in three patients, amyotrophic lateral sclerosis in three patients, Charcot-Marie-Tooth disease in two patients and kyphoscoliosis, multiple sclerosis, and cervical myelopathy in one patient each.

Continue Reading »

04 May

Tracheostomy Ventilation

Patients with traumatic quadriplegia or progressive neuromuscular conditions may require long-term ventilatory support that is most frequently managed by intubation and subsequent tracheostomy and tra­cheostomy intermittent positive pressure ventilation (TIPPV) with an inflated cuff. Patients are often unnecessarily maintained with an inflated cuff longer than necessary. Although cuff deflation and transition to the use of cuffless tracheostomy tubes have been described during ventilator weaning of patients with paralytic respiratory insufficiency, weaning from TIPPV is not always possible. There is often no effort made to optimize the tracheostomy tube diameter and the delivered volumes to permit cuff deflation or removal and there appear to be no guidelines in the literature to facilitate this.

Continue Reading »

03 May

Comparison of Flow-Resistive Work Load due to Humidifying Devices: DISCUSSION part 2

The dE( +) with HMEs was different from that with HWBs. Although dE( +) with HWBs was theoretically zero, it was not actually zero, probably due to delay in expiratory valve opening. The Bear 5 monitors the inspiratory flow rate, and a decrease to one fourth of the peak inspiratory flow rate is the criterion for termination of demand flow and opening of the expi­ratory valve. We selected the square wave flow pattern on the muscle ventilator, a Puritan-Bennett 7200a. Inspiratory flow increased sharply, and after delivery of the set volume, it decreased very sharply. In case of a high inspiratory flow rate, therefore, opening of the expiratory valve on the Bear 5 was delayed and dE( + ) increased. The dE( +) with HMEs was very large compared with that with HWBs because expi­ratory gas must flow across and through the devices. The dE( + )P2 of HMEs was the largest at any inspir­atory flow rate except 30 L/min. This means that, when Paw was monitored at P2, ventilator work done on the mechanical lung was the greatest and a larger volume was delivered to the lung compartment of the mechanical lung. During exhalation, a greater amount of gas flowed across and through the HMEs, resulting in increased dE( + ). The clinical relevance of these AWLs during exhalation was difficult to evaluate. However, as Dodd et al mentioned, the efficiency of expiratory muscles is one half that of inspiratory muscles and they therefore need more oxygen. From this point of view, disposable HMEs might increase the oxygen cost of breathing, especially that of patients with obstructive lung disease. Furthermore, when HMEs become humid after long use, a further in­crease in resistance and AWL may occur.

Continue Reading »

02 May

Comparison of Flow-Resistive Work Load due to Humidifying Devices: DISCUSSION

There are various humidifying devices and their ability to heat and humidify has been evaluated extensively. Unfortunately, all humidifiers exhibit airway flow impedance and this flow resistance against continuous flow also has been evaluated. Their flow impedance sometimes causes malfunctioning of the low pressure alarm which detects a ventilator disconnection. Little attention, however, has been given to the AWL imposed on a patient due to the flow resistance of the device.

Continue Reading »

01 May

Comparison of Flow-Resistive Work Load due to Humidifying Devices: RESULTS

All measurements in the present study showed excellent reproducibility, and deviation for each device under the same conditions was within 3 percent of the coefficient of variance. Standard deviation was very small and for all experimental conditions all values were statistically significant except comparison for different pressure-monitoring sites of Vapoi-Phase.

Thedl(-)

Figure 3 shows dl( —) for each device for four different inspiratory flow rates and Paw monitoring at three different sites. All devices exhibited the largest dl( —) when the pressure monitoring was done at PI. The dl( —) was smallest when the pressure was mon­itored at P2 under inspiratory flow rates of 30 and 60 L/min and all humidifiers had little effect on dl( —). Under a high inspiratory flow rate, dl( —)P3 of the HWB systems was less than dl( —)P2. The reason for this phenomenon was not clear. As the inspiratory flow rate increased, dl(-), especially dl( —)P1, of all devices increased. The cascade humidifier exhibited the largest dl( — )P1 at any inspiratory flow rate. Vapor- Phase (Inspiron) represented almost no resistance to gas flow and its dl( —) was zero. When Paw was monitored at P3, dl( —) for HME was greater than that of HWB.

Continue Reading »

30 Apr

Comparison of Flow-Resistive Work Load due to Humidifying Devices

Endotracheal intubation results in a bypass of the natural warming and humidifying process through the nasopharynx, and a humidifying device is therefore necessary during respiratory tract manage­ment. Although FDA guidelines place some require­ments on the flow-resistance characteristics of humid­ifiers at continuous flow rates, there have been few reports investigating the flow-impedance characteris­tics of humidifiers under dynamic conditions and the changes in respiratory work load due to humidifiers. In the present study, we calculated the amount of AWL imposed by different types of humidifying de­vices and evaluated their flow-impedance character­istics. In addition to this, we examined the ventilator factors which affect the values of AWL.

Continue Reading »

Pages: Prev 1 2 3 4 5 6 7 8 9 10 ...85 86 87 Next

© 2008 HIV/AIDS News & Information